Model Power supply Network frequency Input average power Maximum peak power (welding phase) Medical Device Class Insulation Class IP degree of protection Mode of use Working conditions Temp./RH/Press. Storage conditions Temp./RH/Press. External connections Size Weight |
Implant-Weld 300 230V AC 50-60 Hz 10 W 1 KW 2 B I - Handpiece Type BF IPXO Continuous operation 10ºC - 35ºC / 30%-90% / 700-1100 hPa 5ºC - 50ºC / 30%-90% / 700-1100 hPa Pedal 30x22x17 cm (length x width x height) About 7 kg |
Power Emission mode Time between two pulses Mode of use Welding power stability |
20J – 300J Selectable: Single - Double pulse 5s Continuous ±20% |
Protection against anesthetic: Equipment not suitable for use in the presence of a flammable anesthetic mixture with oxygen or nitrous oxide.